| + | ==== How to connect the IKEM coil ==== | ||
| + | - Ordered List ItemTurn off the gradients (green switch in the middle case in the technical room) | ||
| + | - Ordered List ItemRemove the volume coil if connected and disconnect the red labeled cable | ||
| + | - Connect the orange labeled cable from the RF Interface Adapter | ||
| + | - Ordered List Item Open the electronic case at the MR bore (left case) | ||
| + | * Turn off the black switch (under the green lights) | ||
| + | * Disconnect the first round connector (Power Supply) | ||
| + | * Connect this cable with male cable from the RF Interface Adapter | ||
| + | * Connect female cable from the RF Interface Adapter into the Power Supply | ||
| + | * Turn on the black switch | ||
| + | - Connect the gray cable into the RF Interface Adapter in the Low Power | ||
| + | - Turn on the gradients | ||
| + | - Before measuring, the active coil configuration has to be set (Scan Control -> Scanner -> Change Active Coil Configuration -> IKEMspinal) | ||